Impedance spectroscopy of tripolar concentric ring electrodes with Ten20 and TD246 pastes

Document Type

Conference Proceeding

Date of Original Version

9-13-2017

Abstract

Electrodes are used to transform ionic currents to electrical currents in biological systems. Modeling the electrode-electrolyte interface could help to optimize the performance of the electrode interface to achieve higher signal to noise ratios. There are previous reports of accurate models for single-element biomedical electrodes. In this paper, we measured the impedance on both tripolar concentric ring electrodes and standard cup electrodes by electrochemical impedance spectroscopy (EIS) using both Ten20 and TD246 electrode paste. Furthermore, we applied the model to prove that the model can predict the performance of the electrode-electrolyte interface for tripolar concentric ring electrodes (TCRE) that are used to record brain signals.

Publication Title, e.g., Journal

Proceedings of the Annual International Conference of the IEEE Engineering in Medicine and Biology Society EMBS

Share

COinS